This website uses cookies primarily for visitor analytics. Certain pages will ask you to fill in contact details to receive additional information. On these pages you have the option of having the site log your details for future visits. Indicating you want the site to remember your details will place a cookie on your device. To view our full cookie policy, please click here. You can also view it at any time by going to our Contact Us page.

Helping ensure integrity of life-saving medical device

07 April 2017

Intertronics’ customer BPR Medical is a small British company which combines creativity and entrepreneurialism to design, develop and manufacture world-beating products for export worldwide and use in the United Kingdom. As with any manufacturing company, the success of its products depends to a considerable degree on assembly methods.

BPR Medical Firesafe Cannula Valve

The company’s latest success story is the Bidirectional Firesafe Cannula Valve, used for in-home medical oxygen supplies to extinguish fires which can occur in the line between the concentrator and the user’s mask or nasal cannula. Such fires can occur if the line comes into contact with ambient flame – such as a candle – or more commonly if the patient is a smoker. 

The UK’s Home Oxygen Service supports some 90,000 patients. In the US, accident & emergency units attend to in excess of 1,000 thermal burns a year caused by ignitions associated with home medical oxygen, of which smoking was by far the leading cause.

Life and death integrity

BPR’s innovative inline Bidirectional Firesafe Cannula Valve extinguishes the fire by cutting off the oxygen supply. It is a development of an existing, unidirectional design and acts as a thermal fuse whereby the oxygen supply is cut off when a fusible component softens as a result of the heat from an approaching fire in the oxygen delivery tube. Integrity of operation is vital and could literally represent the difference between life and death. 

When BPR had chosen a two-part epoxy to bond both halves of the valve’s body, the next task was to find a dispensing solution that would assure deposition of a precise, repeatable volume of the adhesive, metered and mixed in the correct ratio, on to a cylindrical assembly. The resulting bond integrity would help ensure conformance for CE marking under European Medical Device Directive.

Technical Director Ben Johnson’s team evaluated a number of possibilities, including premixing and dispensing via a pinch tube valve. This yielded inconsistent results due to changes in viscosity which begin to occur naturally as soon as epoxy is mixed, exacerbated by temperature fluctuations. It also took an unacceptable amount of set-up and clean-up time.

The accuracy of volumetric dispensing

preeflow eco-DUO and eco-CONTROL

Having determined that volumetric technology would provide a better solution, BPR contacted Intertronics and two other vendors to arrange equipment demonstrations. Trials using the selected epoxy were conducted in a wide ambient temperature range, and rigorous pull, flexural and other testing of the assembled valve, revealed that the preeflow eco-DUO precision metering, mixing and dispensing system suggested by Intertronics was the correct solution. It offers ±1 percent dosing accuracy, >99 percent repeatability and can dispense volume flows of 0.2 to 32ml per minute, with a minimum volume of 0.01ml. The preeflow positive displacement technology means that the volume dispensed is not affected by viscosity changes in the material. 

Ben Johnson and Product Development Program Manager Mike Brudenell worked with Intertronics to configure the dispenser into its bespoke manufacturing jig, into which the two populated halves of the Firesafe valve body are manually loaded prior to being rotated whilst the epoxy is applied. Each of the 20,000 units per month produced at the time of writing is non-destructively tested, whilst regular samples are subject to a three-point flexural test. 

The system operates an automatic purge at the end of the shift, eliminating the need for cleaning down. Control is via a preeflow plug ‘n’ mix interface and stored settings ensure consistency and repeatability, and eliminate daily set-up time. Thanks to preeflow’s accuracy, the volume of epoxy applied has been reduced to 0.05g from the pinch tube valve’s 0.06g.

Ben Johnson states that they had to go through a number of stringent processes to ensure the integrity of the bonding of the two body halves. An essential part of this was how the two-part epoxy adhesive was applied: to ensure a highly accurate, repeatable dose on a rotating jig, but also to maintain the correct dispense volume regardless of viscosity changes caused by temperature. 

Ben estimates the return on investment to be in the region of £2,000 a month.

See accompanying video: https://youtu.be/_xkT2DNxCAo